Biotage® Extrahera™ HV-5000
The Biotage® Extrahera™ HV-5000 is a state-of-the-art automated sample preparation workstation specifically designed for high-volume biological, pharmaceutical, and environmental sample processing. Engineered for laboratories performing solid phase extraction (SPE) and supported liquid extraction (SLE) workflows, this system accommodates classical 3, 6, 10, and 15 mL column formats for maximum flexibility. With an advanced 4-channel pipette head capable of handling 250–5000 µL per channel, the HV-5000 ensures precise and reproducible liquid handling for demanding high-volume extractions. A compact footprint and integrated touchscreen interface make it easy to use in a variety of lab environments—including benchtops, fume hoods, or mobile labs.
- High-Volume Pipetting — A 4-channel pipette head aspirates and dispenses 250–5000 µL per channel for large-volume sample preparation.
- Supports Multiple Column Formats — Compatible with 3, 6, 10, and 15 mL SPE and SLE columns for full workflow flexibility.
- Compact & Portable — Small footprint with integrated touchscreen—ideal for benchtop, fume hood, or mobile lab setups.
- SMART Pipette Loading — Prevents air aspiration and sample overflow for consistent and safe sample handling.
- Positive Pressure Processing — Ensures uniform flow across all samples for reproducible SPE and SLE extractions.
- Three-Waste Stream Management — Independently control aqueous, organic, and chlorinated waste lines to reduce disposal costs and environmental impact.
- User-Friendly Software — Intuitive interface requires minimal training; perfect for both novice and expert users.
- GLP-Compliant Operation — Designed for regulated laboratory environments needing reliable, traceable sample preparation.
- Pharmaceutical & Biotech Sample Preparation
- Forensic and Toxicology Analysis
- Clinical Research Sample Processing
- Environmental Pollutant Extraction
- Food and Beverage Safety Testing
| Category | Specification |
|---|---|
| Liquid Handling | |
| Pipette Pump Capacity | 250–5000 μL |
| Pipetting Specifications | For 5000 μL pipette tip (P/N 417007): • 250 μL: ± 2.0% accuracy, 3.0% CV • 500 μL: ± 1.5% accuracy, 2.0% CV • 1000 μL: ± 1.0% accuracy, 1.5% CV • 5000 μL: ± 0.5% accuracy, 1.0% CV |
| Solvent Capacity | 5 solvent pumps with automatic feed from external reservoirs + 5 manually filled reservoirs (5 × 20 mL, 5 × 40 mL, 5 × 100 mL) |
| Pressure Unit | Positive Pressure Unit: • Processing: 0–5 Bar (0.1 Bar increments) • Plate Dry: 5 Bar at 600 mL/min |
| Sample Processor | |
| Format | • 48-position racks for 3 mL columns • 24-position racks for 3 & 6 mL columns • 12-position racks for 10 mL columns • 12-position racks for 15 mL columns |
| Plate Positions | • 6 upper deck positions (samples, extraction plates, pipette tips, solvents) • 4 lower carousel positions (multiple elution & waste) |
| System Requirements | |
| Temperature | Operating: 18–32 °C Storage/Transport: -25–60 °C |
| Humidity | 10–90% RH |
| Power Supply | 100–240 V, 50/60 Hz |
| Vacuum Supply | Partial pressure: < 500 mbar (50 kPa; 7 psi) Flow rate: ≥ 8.3 L/min (0.29 cfm) |
| Physical Characteristics | |
| Weight | 80 kg (176 lbs) |
| Dimensions (W × D × H) | Without touch screen: 610 × 510 × 790 mm (24.0” × 22.1” × 31.1”) With touch screen: 860 × 570 × 790 mm (33.8” × 22.4” × 31.1”) |
| Max. Sound Level | 65 dB(A) |