Home / Products / Laboratory Equipment / Aerosol Sciences / Particles / Condensation Particle Counter 3750

Condensation Particle Counter 3750

The Condensation Particle Counter (CPC) 3750 is a next-generation nanoparticle counter designed for precision measurements down to 7 nm (D50). Using butanol as the working fluid and operating at a consistent 1 L/min flow rate, the CPC 3750 delivers highly reliable performance across research and industrial applications. It seamlessly integrates with SMPS™ configurations, making it ideal for real-time aerosol studies, filter testing, and atmospheric research.

Contact Us for a Quote
Features
  • Detection down to 7 nm (D50) for sub-10 nm particle visibility
  • Extended concentration range up to 100,000 particles/cm³
  • High data rate (50 Hz) for dynamic aerosol behavior analysis
  • Reliable internal memory with data storage for continuous monitoring
  • Pulse Height Monitoring (PHM) for enhanced diagnostics
  • Water-removal system ensures performance in humid environments
  • Rapid response time (T10-90 < 1 sec) for fast concentration changes
  • Flow protection shutoff when inlet is blocked
  • Upgradeable to 1 nm CPC with Nano Enhancer compatibility
  • Software-optional operation with built-in data logging
Applications
  • Nanoparticle and aerosol research
  • Filter efficiency testing
  • Combustion and engine emission studies
  • Environmental and atmospheric studies
  • SMPS™ system integration for detailed particle size analysis
  • Health impact assessments of ultrafine particles
  • Environmental science and air quality monitoring
  • Industrial hygiene and occupational safety
  • Combustion science and automotive R&D
Specifications
Parameter Specification
Particle Size Range 7 nm minimum detectable (D50), verified with monodisperse sucrose particles; >3 µm maximum detectable particle size
Particle Concentration Range Up to 100,000 (1×10⁵) particles/cm³; single particle counting with continuous live-time coincidence correction
Particle Concentration Accuracy ±5% at <100,000 particles/cm³
False Background Counts <0.001 particles/cm³ (12-hour average)
Response Time <1 s for 90%→10% (T10-90, T90-10); ~2 s for 0→95% (T95)
Flow System 1.0 ± 0.05 L/min inlet & counting flow (volumetric); requires external vacuum source ≥60 kPa (18 in Hg); compatible pumps: Model 3032, 3032-EC
Liquid System Butanol (n-Butyl alcohol, not included) as working fluid; internal water removal pump for condensate (useful in humid environments)
Ambient Operating Conditions Temperature: 10–35 °C (50–95 °F); Humidity: 0–90% RH, non-condensing; Pressure: 75–105 kPa (0.75–1.05 atm)
Electrical 100–240 VAC, 50/60 Hz, 200 W max
Data Storage Internal memory: approx. 1 year at 50 Hz data rate
Weight ~6.6 kg (14.6 lbs)
Dimensions (L × W × H) 27.5 × 18.3 × 29.9 cm (10.83 × 7.21 × 11.76 in), excluding fill bottle & bracket