Home / Products / Laboratory Equipment / Aerosol Sciences / Ozone Monitoring / Model 211 Scrubberless Ozone Monitor

Model 211 Scrubberless Ozone Monitor

The Model 211 Scrubberless Ozone Monitor by 2B Technologies is the world’s most accurate ozone monitor for regulatory and research-grade air quality monitoring. Featuring dual-beam UV absorption, an extended 30 cm optical path, and a patented gas-phase scrubber using NO titration, the Model 211 delivers unmatched precision (<0.5 ppb) with zero interference from VOCs, mercury, or particulate matter. Certified by the U.S. EPA as a Federal Equivalent Method (FEM) (EQOA–0514–215), it is the ideal tool for air quality agencies and environmental researchers in non-attainment regions requiring data accuracy and compliance confidence. The unique scrubberless design ensures that even in highly polluted environments, your ozone readings are reliable, interference-free, and backed by innovative nitric oxide (NO) titration technology.

Features
  • Dual-beam 30 cm optical bench with <0.5 ppb precision
  • EPA-certified Federal Equivalent Method (FEM): EQOA–0514–215
  • Patented scrubberless gas-phase NO titration technology (US Patent No. 8,395,776 B2)
  • Interference-free measurement of ambient ozone, even in polluted environments
  • DewLine™ system for water vapor interference elimination
  • SD flash memory for virtually unlimited data storage
  • Internal long-life sample pump (20,000 hours)
  • Real-time clock for accurate time-stamping
  • Convenient user interface with RS-232 remote operation
  • Internal nitric oxide (NO) generator using non-toxic N₂O cartridges
  • Compatible with 8g or 16g “Whippit”-type cartridges for 4–8 hours of NO supply
  • Low power consumption and compact design
  • Ideal for regulatory compliance, research, and urban air quality monitoring
Applications
  • Ambient Ozone Monitoring in regulatory and research settings
  • Compliance Monitoring in non-attainment areas under the Clean Air Act
  • Air Pollution Studies with high VOC, PM, or mercury background
  • Reference Data Collection for calibration and QA/QC programs
  • Mobile Air Quality Monitoring Units
  • Remote and Urban Monitoring Stations
Specifications
Parameter Specification
Measurement Principle UV Absorbance at 254 nm; Dual Beam
Certifications Federal Equivalent Method (FEM), 0–500 ppb for 20–30 °C: EQOA–0514–215; CE
Ozone Scrubber Technology Gas-phase scrubber using NO, provided directly or by photolysis of N₂O (supplied by compressed gas cylinder or portable liquid N₂O cartridge)
Linear Dynamic Range 0–2,000 ppb (0–2 ppm)
Resolution 0.1 ppb
Precision (1σ rms noise) Greater of 0.5 ppb or 1% of reading (10-s average)
Accuracy Greater of 1.0 ppb or 2% of reading
Limit of Detection (2σ) 1.0 ppb for 10-s average
NIST-Traceable Calibration Yes
Flow Rate (nominal) ~2 L/min
Flow Rate Requirement >1.2 L/min
Baseline Drift <1 ppb/day; <3 ppb/year
Sensitivity Drift <0.5%/day; <3%/year
Response Time (100% step change) 20 s (10-s averaging); 4 s (2-s measurement)
Measurement & Averaging Times 2 s, 10 s, 1 min, 5 min, 1 hr
Internal Data Logger Capacity 16,383 lines (10-s avg. = 1.9 days; 1-min avg. = 11 days; 5-min avg. = 1.9 mo; 1-hr avg. = 1.9 yr)
SD Card Logger Capacity ≥2 GB (>5 years capacity at 10-s measurement mode)
Ozone Units ppb, ppm, pphm, µg m⁻³, mg m⁻³
Pressure Units mbar, torr
Temperature Units °C, K
T and P Corrected Yes
Operating Temperature Range 10 to 50 °C
Operating Altitude Range ~0–13.5 km (150–1,013 mbar); optional ~0–25 km (30–1,013 mbar) with upgraded pressure sensor
Power Requirement 11–14 V DC or 120/240 V AC, 1 A at 12 V, 12 W
Size Rackmount: 43 × 37 × 14 cm (17″ × 14.5″ × 5.5″)
Weight 6.7 kg (14.7 lb)
Data Outputs RS232, SD data card, LCD, 0–2.5 V Analog output
Data Transfer Baud Rates 2400, 4800, 19200
Output Ranges User-defined scaling factor in menu
Special Features DewLine™ humidity control, Long-life pump (20,000 hr), Flow meter
Options External particle filter, Bluetooth, Serial-to-USB adapter, Lamp heater, High-altitude upgrade (0–25 km pressure sensor)