SMPS 3938W50-CEN10 Wide-Range Ambient Particle Sizer
The TSI SMPS™ 3938W50-CEN10 is a wide-range Scanning Mobility Particle Sizer designed for continuous ultrafine particle monitoring in ambient air. Fully compliant with CEN/TS 17434:2020, this high-precision instrument meets European standards for atmospheric nanoparticle sizing. With a particle size range from 10 nm to 800 nm, advanced data logging capabilities, and integration with monitoring networks, it is the ideal solution for urban air quality stations, environmental research, and regulatory compliance. Equipped with the advanced DMA 3083 (based on the TROPOS Vienna-type DMA) and ECAC-accredited calibration, the 3938W50-CEN10 delivers standardized, traceable nanoparticle measurements with minimal downtime. Its robust design allows 24/7/365 operation—providing accurate, harmonized data for agencies, scientists, and environmental professionals monitoring particle pollution.
- Compliant with CEN/TS 17434:2020 for standardized ultrafine particle size monitoring
- Wide size range: 10 nm to 800 nm using the new DMA 3083
- Scan time as fast as 1 minute for capturing dynamic particle events (e.g., traffic, airport emissions)
- 24/7/365 continuous monitoring for long-term ambient measurement
- Factory-calibrated by an ECAC-accredited facility (for model 3938W50-CEN10)
- Component modularity: upgrade or modify size range with interchangeable SMPS components
- Advanced sampling system with loss correction, auto-export, and power-failure recovery
- Raw and corrected data export for harmonized network analysis
- Seamless integration with relative humidity and temperature sensor (RHT3000)
- Reliable operation in environmental monitoring networks and research setups
- Optionally available as 3938W50 model (without ECAC calibration)
- Regulatory ambient air quality monitoring
- Environmental chamber aerosol studies
- Indoor air quality investigations in urban and industrial settings
- Health impact research of ultrafine particles
- Basic and applied aerosol research
- Networked air monitoring stations (e.g., GAW, EPA, EU-wide networks)
| Parameter | Value |
|---|---|
| Aerosol Flow Rate | 1 L/min |
| Sheath Flow Rate | 2 to 15 L/min, user-selectable |
| Particle Size Range | 10 to 800 nm |
| Measurement Time | 1 to 10 minutes, user-selectable |
| Working Fluid for CPC | n-butyl alcohol (butanol) |
| Ambient Operating Conditions | Temperature: 10–35 °C; Pressure: 75–105 kPa; Humidity: 0–90% (non-condensing). Temperature and pressure affect available particle size range. |
| Data Acquisition | Continuous with PC-based software. Optional monitoring module enables automatic export of raw/final concentrations, auto-recovery after power outage, and correction for sampling losses. |
| Aerosol Neutralizer Options (ordered separately) | 3077A: 370 MBq (10 mCi) Kr-85, half-life 10.8 years; 3088: Soft X-ray <9.5 keV, ~8,760 operating hours; 6005931: Lead shielding column for 3077/3077A inside 3082 classifier |
| Communication & User Interfaces | Ethernet (8-wire RJ-45, 10/100 BASE-T, TCP/IP), configurable for DHCP or manual; RS-232 CPC-to-Classifier; embedded touch display for local diagnostics |
| Power Requirements | 3750 CPC: 200 W; 3082: 200 W |
| Dimensions (H × W × D) / Weight | 3082: 40 × 28 × 40 cm / 14.2 kg; 3083: 47 × 13.2 × 15.9 cm / 8.5 kg; 3750: 27.5 × 18.3 × 29.9 cm / 6.6 kg; Assembled 3938W50-CEN SMPS system: 67.5 × 40.5 × 40 cm / 29.3 kg* |